1. Introducción
This manual provides essential information for the safe and effective operation of the UEi IRT803 Digital Insulation Resistance Tester. The IRT803 is a versatile instrument designed for measuring insulation resistance, AC/DC voltage, and continuity. It features a 1000V AC/DC measurement capability, auto-ranging, a backlit display, and a DMM rotary style selector. Key functions include Lock, Compare, and Polarization Index (PI)/Dielectric Absorption Ratio (DAR) tests. Please read this manual thoroughly before use to ensure proper handling and to maximize the instrument's performance.
2. Información de seguridad
Always adhere to safety precautions when using electrical testing equipment. The IRT803 is designed with safety in mind, featuring a CATIV rating up to 600V and CATII up to 1000V, ensuring protection in various jobsite environments. Failure to follow safety instructions can result in electric shock, fire, or damage to the instrument.
- No utilice el comprobador si parece estar dañado o si los cables de prueba están dañados.
- Asegúrese de seleccionar la función y el rango correctos antes de realizar cualquier medición.
- Always disconnect power to the circuit under test before connecting or disconnecting test leads for resistance measurements.
- Utilice equipo de protección personal (EPP) adecuado, como guantes aislantes y gafas de seguridad.
- Do not operate the tester in wet conditions or in the presence of explosive gases or dust.
- Reemplace las baterías rápidamente cuando aparezca el indicador de batería baja para garantizar lecturas precisas.
3. ¿Qué hay en la caja?
Al desembalar, verifique que todos los elementos enumerados a continuación estén presentes y sin daños:
- UEi IRT803 Insulation Resistance Tester
- Silicone test leads with alligator clips (red and black) (ATL57)
- Insulation test probe with alligator clips (red & black)
- 4 pilas AA
- Manual de usuario
- Estuche blando

Image 3.1: All components included with the UEi IRT803 Insulation Resistance Tester, including the meter, test leads, probes, batteries, manual, and soft carrying case.
4. Producto terminadoview
The IRT803 is designed for ease of use and durability. It features an easy-to-read LCD display with a backlight for visibility in various lighting conditions, and a work light to illuminate dark environments. The unit is protected by a rugged rubber boot, which also includes convenient test lead holders for compact storage. An auto power-off function helps conserve battery life.
4.1. Características y funciones principales
- 600V AC/DC Measurement: Capable of measuring both alternating and direct current voltages hasta 600V.
- Resistencia de aislamiento: Measures resistance from 0.01MΩ up to 2GΩ with test voltages of 50V, 100V, 250V, 500V, and 1000V.
- Continuidad: Para comprobar la continuidad del circuito.
- Compare Reading Function: Allows for comparison against a set reference value.
- Prueba PI/DAR: Automatic calculation of Polarization Index and Dielectric Absorption Ratio for advanced insulation diagnostics.
- A-Hold/Hold: Congela la lectura mostrada para una grabación conveniente.
- Lock Reading: Locks the test voltage for continuous insulation testing.
- Rotary Dial Selector: Intuitive selection of measurement functions and test voltages.
- Bota protectora de goma: Enhances durability and provides lead storage.

Imagen 4.1: Frente view of the IRT803 showing the LCD display, rotary selector, and function buttons.

Image 4.2: Visual representation highlighting the main functions and features of the IRT803.
5. Configuración
5.1. Instalación de la batería
The IRT803 requires four (4) AA batteries for operation. To install or replace batteries:
- Asegúrese de que el comprobador esté apagado.
- Localice el compartimiento de la batería en la parte trasera de la unidad.
- Utilice un destornillador para abrir la tapa del compartimento de la batería.
- Insert the four AA batteries, observing the correct polarity (+/-) as indicated inside the compartment.
- Cierre bien la tapa del compartimento de la batería.
5.2. Conexión de los cables de prueba
Connect the test leads to the appropriate input jacks on the tester. For most measurements, the red lead connects to the 'V Ω INSULATION' jack and the black lead connects to the 'COM' (common) jack. Ensure connections are firm and secure.

Image 5.1: The IRT803 with test leads properly connected to the input jacks.
6. Instrucciones de funcionamiento
6.1. Powering On/Off and Display Backlight
- To power on the unit, rotate the selector dial from 'OFF' to any desired measurement function.
- To power off, rotate the selector dial back to 'OFF'. The unit also features auto power-off to save battery life after a period of inactivity.
- Presione el SELECCIONAR button (often indicated by a light bulb icon) to toggle the display backlight on or off.
6.2. Prueba de resistencia de aislamiento
The IRT803 can perform insulation resistance tests at various voltages.
- Seleccionar volumen de pruebatage: Rotate the selector dial to the desired insulation test voltage (50 V, 100 V, 250 V, 500 V o 1000 V).
- Conectar los cables de prueba: Connect the red test lead to the circuit under test and the black test lead to the ground or common point. Ensure the circuit is de-energized.
- Iniciar prueba: Press and hold the large red PRUEBA button to begin the insulation test. The test voltage will be applied, and the resistance reading will be displayed.
- Función de bloqueo: For continuous testing, press the CERRAR button after initiating the test. This will maintain the test voltage without needing to hold the PRUEBA botón. prensa PRUEBA de nuevo para desbloquear.
- Compare Function: Utilice el COMPARAR button to set a reference value and compare subsequent readings against it.
- Prueba PI/DAR: Select the PI/DAR function on the dial. The meter will automatically perform the Polarization Index and Dielectric Absorption Ratio tests, displaying the calculated values.
- Read and Record: Observe the reading on the LCD. Use the SOSTENER button to freeze the display for easier recording.
- Descargar: After completing an insulation test, the circuit under test should be discharged. The IRT803 typically handles this automatically, but always exercise caution.

Image 6.1: The IRT803 being used to measure insulation resistance on electrical wiring.

Imagen 6.2: Un primer plano view of the test leads securely connected to the IRT803's input terminals.
6.3. Vol.tagPruebas electrónicas (CA/CC)
Para medir CA o CC voltage up to 600V:
- Rotate the selector dial to the 'V' position.
- Connect the red test lead to the positive or live point and the black test lead to the negative or neutral/ground point of the circuit.
- La pantalla mostrará el volumen medido.tage. The meter will automatically detect AC or DC voltage.
6.4. Pruebas de continuidad
To check for continuity in a circuit:
- Rotate the selector dial to the 'Ω' (Ohms) position, which typically includes continuity.
- Connect the test leads across the component or circuit path you wish to test.
- If there is continuity (low resistance), the meter will typically emit an audible beep and display a low resistance value.
7. Mantenimiento
7.1. Limpieza
To maintain the accuracy and longevity of your IRT803:
- Limpia la casing con anuncioamp paño y detergente suave. No utilice abrasivos ni disolventes.
- Ensure the unit is completely dry before storage or next use.
7.2. Reemplazo de la batería
Replace batteries as described in Section 5.1 when the low battery indicator appears on the display. Always use fresh AA alkaline batteries.
7.3. Almacenamiento
When not in use, store the IRT803 in its soft carrying case in a cool, dry environment, away from direct sunlight and extreme temperatures. If storing for extended periods, remove the batteries to prevent leakage.
8. Solución de problemas
If you encounter issues with your IRT803, refer to the following common problems and solutions:
| Problema | Posible causa | Solución |
|---|---|---|
| El probador no se enciende | Baterías agotadas o mal instaladas | Verifique la polaridad de la batería; reemplácela con baterías AA nuevas. |
| No reading on display | Incorrect function selected; open circuit; poor test lead connection | Verify selector dial position; ensure test leads are firmly connected to the circuit and the meter; check for breaks in the circuit. |
| Lecturas inexactas | Low battery; environmental interference; incorrect range/function | Replace batteries; move away from strong electromagnetic fields; ensure correct function and test voltage are selected. |
| Test leads not working | Damaged leads; poor connection | Inspect leads for damage; ensure they are fully inserted into the jacks. Replace if damaged. |
If the problem persists after attempting these solutions, please contact UEi Test Instruments customer support.
9. Especificaciones
| Parámetro | Valor |
|---|---|
| Número de modelo | IRT803 |
| Marca | Instrumentos de prueba UEi |
| Prueba de aislamiento vol.tages | 50 V, 100 V, 250 V, 500 V, 1000 V |
| Rango de resistencia de aislamiento | 0.01MΩ to 2GΩ |
| Insulation Resistance Accuracy | ±1.5-3% + 5 digits (0.01MΩ-500MΩ) |
| VolumentagMedición electrónica | Hasta 600 V CA/CC |
| Rango de continuidad | 20.00 Ω to 20.0 kΩ (as per product description) |
| Clasificación de seguridad | CATIV 600V, CATII 1000V |
| Fuente de poder | 4 pilas AA |
| Mostrar | LCD con luz de fondo |
| Dimensiones del producto | 2.7 x 3.52 x 8.27 pulgadas |
| Peso del artículo | 1.8 libras |
| Color | Amarillo |
10. Información de garantía
The UEi IRT803 Insulation Resistance Tester is backed by a Garantía limitada de 1 años from the date of purchase. This warranty covers defects in materials and workmanship under normal use. It does not cover damage resulting from misuse, unauthorized modification, accident, or neglect. For specific terms and conditions, please refer to the warranty card included with your product or contact UEi Test Instruments customer support.

Imagen 10.1: Example of a Certificate of Conformity, indicating adherence to quality standards.
11. Soporte
For technical assistance, troubleshooting beyond this manual, or warranty claims, please contact UEi Test Instruments customer support. You can find more information and contact details by visiting the official UEi Test Instruments websitio o su página de tienda de Amazon.
UEi Test Instruments Store: Visita la tienda
Documentos relacionados - IRT803
![]() |
UEi DL200 CATIII Clamp Medidor y vol.tage Manual de instrucciones del probador Manual de instrucciones completo para el UEi DL200 CATIII Clamp Medidor y vol.tagProbador electrónico, que detalla características, especificaciones, advertencias de seguridad y procedimientos de medición. |
![]() |
UEi DL599 Inalámbrico TRMS Clamp Medidor con pruebas de motores trifásicos y desequilibrados User manual for the UEi DL599 Wireless TRMS Clamp Meter, detailing its features for 3-phase and unbalanced motor tests, electrical measurements, and safe operation. Includes specifications, functions, and troubleshooting. |
![]() |
UEi DL389B TRMS Pantalla doble Clamp Manual de instrucciones del medidor Manual de instrucciones completo para el monitor de doble pantalla UEi DL389B TRMSamp Medidor que cubre funciones, características, especificaciones, advertencias de seguridad e instrucciones de funcionamiento para varias mediciones, incluido el volumen.tage, corriente, resistencia, capacitancia, temperatura, frecuencia y vol. sin contactotage detección. |
![]() |
UEi DL599 Inalámbrico TRMS Clamp Meter: Features and Specifications Explore the UEi DL599, a versatile wireless TRMS clamp meter with advanced 3-phase and unbalanced motor testing capabilities. Discover its key features, comparison with other DL500 series models, and technical specifications. |
![]() |
Instrumentos de prueba de refrigeración HVAC/R UEi: SSM1, RLD, DMG, DRS, HA1 integral sobreview de las herramientas de HVAC/R y refrigeración de UEi Test Instruments, incluido el medidor de supercalentamiento/subenfriamiento SSM1, los detectores de fugas RLD10/RLD15, los medidores de micras DMG100/DMG200, la báscula DRS220 y el módulo DRSCM, el analizador de compresores HA1 y otros componentes relacionados.amp Medidores y comprobadores de temperatura. |
![]() |
UEi Test Instruments Catalog: Comprehensive Range of Measurement Tools Explore the latest UEi Test Instruments catalog featuring a wide array of digital multimeters, clamp meters, combustion analyzers, thermometers, and other essential testing equipment for professionals. |
Ask a question about this manual
Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.