MAXUS D214

MAXUS Digital Milligram Scale

Model: D214 - User Instruction Manual

Introducción

Thank you for choosing the MAXUS Digital Milligram Scale. This precision instrument is designed for highly accurate measurements up to 50g with a readability of 0.001g, making it ideal for a wide range of applications including powders, jewelry, and other delicate items. Please read this manual thoroughly before operation to ensure proper use and to maximize the lifespan of your scale.

Contenido del paquete

Verifique que todos los artículos enumerados a continuación estén incluidos en su paquete:

  • MAXUS Digital Milligram Scale (Model D214)
  • Peso de calibración de 50 g.
  • Powder Trickler
  • Plato de pesaje
  • Pinzas
  • Cuchara
  • Cable USB
  • Pilas AAA (2)
MAXUS Digital Milligram Scale and included accessories

Image: MAXUS Digital Milligram Scale with all included accessories, including the scale, calibration weight, powder trickler, weighing pan, tweezers, scoop, USB cable, and batteries.

Configuración

  1. Fuente de alimentación: The scale can be powered by 2 AAA batteries (included) or via the included USB cable. Insert batteries into the compartment on the underside of the scale, or connect the USB cable to a power source.
  2. Colocación: Place the scale on a stable, level, and vibration-free surface. Avoid areas with drafts, extreme temperature changes, or direct sunlight, as these can affect accuracy.
  3. Encendido inicial: Presione el ENCENDIDO/APAGADO Botón para encender la báscula. La pantalla mostrará "88888" y luego "0.000g".
MAXUS Digital Milligram Scale power options

Image: Illustration showing the two power supply options for the MAXUS Digital Milligram Scale: Battery Mode and USB Mode.

Instrucciones de funcionamiento

  1. Encendido/apagado: Presione el ENCENDIDO/APAGADO Botón para encender o apagar la báscula.
  2. Conversión de unidades: Presione el M (Mode) button to cycle through the available weighing units: g (grams), oz (ounces), ozt (troy ounces), dwt (pennyweights), ct (carats), and gn (grains).
    MAXUS Digital Milligram Scale unit conversion modes

    Image: Diagram illustrating the six different weighing units available on the MAXUS Digital Milligram Scale: g, oz, ozt, dwt, ct, and gn, changed by pressing the 'M' button.

  3. Función de tara: Use the Tare function to subtract the weight of a container, allowing you to measure only the contents.
    1. Encienda la báscula y espere hasta que se estabilice en “0.000g”.
    2. Place the empty container on the weighing pan. The display will show the container's weight.
    3. Presione el T Botón (Tara). La pantalla se reiniciará a "0.000 g".
    4. Add the item(s) to be weighed into the container. The display will show the net weight of the items.
    MAXUS Digital Milligram Scale tare function steps

    Image: Four-step visual guide demonstrating the Tare Function: 1. Open product and press ON/OFF. 2. Put item on weighing pan. 3. Press 'T' key to reset reading to zero. 4. Get the weight of the added substance.

  4. Peso: Place the item directly onto the weighing pan or into a tared container. Wait for the reading to stabilize.
  5. Apagado automático: The scale features an auto shut-off function to conserve battery power. It will automatically turn off after a period of inactivity.

Calibración

Regular calibration ensures the highest accuracy of your scale. Use the included 50g calibration weight for this process.

  1. Ensure the scale is on a stable surface and turned off.
  2. Mantenga pulsado el M (Mode) button. While holding, press the ENCENDIDO/APAGADO button. Release both buttons when the display shows "CAL".
  3. The display will then flash "50.000g".
  4. Carefully place the 50g calibration weight in the center of the weighing pan.
  5. Wait for a few seconds. The display will show "PASS" and then return to "0.000g". Calibration is complete.
MAXUS Digital Milligram Scale calibration steps

Image: Four-step visual guide demonstrating the Calibration process: 1. Press and hold 'M' key until display reads "CAL". 2. Press 'M' key again, display shows "50g". 3. Place 50g weight on platform. 4. LCD displays "PASS".

Mantenimiento

  • Limpieza: Limpie la báscula con un paño suave yamp Paño. No utilice limpiadores abrasivos ni sumerja la báscula en agua.
  • Almacenamiento: Store the scale in a clean, dry environment away from extreme temperatures, humidity, and vibrations. Keep the protective cover closed when not in use.
  • Reemplazo de batería: If the battery indicator appears low or the scale does not power on, replace the AAA batteries.

Solución de problemas

ProblemaPosible causaSolución
La báscula no enciendeLow batteries; Incorrect battery installation; No USB power.Replace batteries; Check battery polarity; Ensure USB cable is properly connected to a power source.
Lecturas inexactas o inestablesUnstable surface; Drafts or vibrations; Needs calibration; Overload; Temperature fluctuations.Place on a firm, level surface; Avoid drafts; Perform calibration; Do not exceed 50g capacity; Allow scale to acclimate to room temperature.
Se muestra "O-Ld" o "EEEE"Sobrecarga.Remove the item immediately. The scale's capacity is 50g.
Se muestra "UNST"Entorno inestable.Ensure the scale is on a stable, vibration-free surface.

Presupuesto

  • Marca: MAXUS
  • Modelo: D214
  • Capacidad: 50 gramos
  • Exactitud: 0.001 gramos
  • Unidades de pesaje: g, oz, ozt, dwt, ct, gn
  • Tipo de pantalla: LCD con pantalla retroiluminada
  • Características especiales: Apagado automático, función de tara
  • Opciones de energía: 2 x AAA Batteries or USB Cable
  • Material: Acrilonitrilo Butadieno Estireno, Acero Inoxidable
  • Dimensiones del producto: 5.43 x 3.15 x 0.98 pulgadas
  • Peso del artículo: 11.2 onzas
MAXUS Digital Milligram Scale dimensions

Image: Diagram showing the dimensions of the MAXUS Digital Milligram Scale: 5.43 inches length, 3.15 inches width, and 0.98 inches height.

Garantía y soporte