Corning PC-420D

Manual de usuario

Corning PC-420D Stirring Hot Plate with Digital Display

Modelo: 6795-420D

For laboratory and scientific applications

Producto terminadoview

The Corning PC-420D Stirring Hot Plate is a precision instrument designed for heating and stirring various liquids and solutions in laboratory environments. It features a durable Pyroceram top plate, digital displays for both stirring speed (RPM) and temperature (degrees Celsius), and intuitive control knobs for precise adjustments. This unit is suitable for a wide range of applications in chemistry, biology, and other scientific fields.

Corning PC-420D Stirring Hot Plate with digital display

Una sobrecarga view of the Corning PC-420D Stirring Hot Plate. The unit features a white Pyroceram top plate, a grey base with two black control knobs (Stir and Heat), and two red LED digital displays for RPM and temperature. A power cord with a three-prong plug is visible on the right side.

Información de seguridad

Please read and understand all safety precautions before operating the Corning PC-420D Stirring Hot Plate. Failure to follow these instructions may result in injury or damage to the equipment.

Configuración

  1. Desembalaje: Carefully remove the hot plate from its packaging. Inspect the unit for any signs of damage during transit. Retain packaging for future storage or transport.
  2. Colocación: Position the hot plate on a clean, stable, and level laboratory bench or work surface. Ensure there is sufficient clearance around the unit for ventilation and safe operation.
  3. Conexión de energía: Connect the three-prong power cord to a grounded 120V/60Hz electrical outlet. The unit is designed for U.S. power standards.
  4. Rod Mount Support: If using a support rod (sold separately), attach it to the built-in rod mount support on the rear of the unit.

Instrucciones de funcionamiento

The Corning PC-420D features intuitive controls for both heating and stirring functions.

Controles sobreview:

Procedimiento:

  1. Encendido: Ensure the unit is plugged in. The power indicator light will illuminate.
  2. Setting Stirring Speed: Place a stir bar into your vessel and position the vessel on the center of the Pyroceram top plate. Turn the "Stir" knob clockwise to the desired RPM. The digital display will show the set speed. Adjust as needed for proper mixing.
  3. Temperatura de ajuste: Turn the "Heat" knob clockwise to the desired temperature in degrees Celsius. The digital display will show the set temperature and will blink until the set temperature is achieved. The temperature can be set in 5-degree C increments, ranging from 5°C to 550°C.
  4. Escucha: Continuously monitor the process. The microprocessor-based control maintains consistent and repeatable temperature settings within ±2°C.
  5. Apagado: When finished, turn both the "Stir" and "Heat" knobs counter-clockwise to the "Off" position. Unplug the unit from the power source when not in use. Allow the hot plate to cool down completely before cleaning or moving. The "HOT TOP" indicator will remain lit until the surface temperature is below 60°C.
  6. Control de temperatura externa: The unit can be used with an external temperature controller and support rod (sold separately) for enhanced temperature precision. Refer to the external controller's manual for specific instructions.

Mantenimiento

Proper maintenance ensures the longevity and optimal performance of your Corning PC-420D Stirring Hot Plate.

Solución de problemas

This section addresses common issues you might encounter with your hot plate.

Problema Posible causa Solución
La unidad no se enciende. No power to outlet; power cord not fully plugged in; unit malfunction. Check power outlet with another device. Ensure power cord is securely connected. If problem persists, contact customer support.
La temperatura no alcanza el punto establecido o fluctúa. Vessel too large for hot plate; insufficient power; external factors (drafts); unit malfunction. Ensure vessel size is appropriate. Minimize drafts. Verify power supply. If issue continues, contact customer support.
Stirrer not working or inconsistent. Stir bar not present or incorrect size; vessel not centered; liquid too viscous; unit malfunction. Ensure a suitable stir bar is in the vessel. Center the vessel on the plate. Reduce liquid viscosity if possible. Contact customer support if problem persists.
Digital display is blinking. Temperature has not yet reached the set point. This is normal operation. The blinking will stop once the set temperature is achieved.

Para problemas que no se enumeran aquí o si los pasos de solución de problemas no resuelven el problema, comuníquese con el servicio de atención al cliente de Corning.

Presupuesto

Característica Detalle
Número de modelo PC-420D (6795-420D)
Material de la placa Vitrocerámica pirocerámica
Dimensiones de la placa 5" x 7" (aprox. 12.7 cm x 17.8 cm)
Rango de temperatura De 5 °C a 550 °C
Precisión de temperatura ±2 °C
Incrementos de temperatura 5°C
Rango de velocidad de agitación Variable (±5% of set speed)
Fuente de poder 120 V/60 Hz
Ampanhelo 5.9 A
Vatios 698 W
Dimensiones del producto (L x An x Al) 7" ​​de profundidad x 5" de ancho x 5" de alto (aprox. 17.8 cm x 12.7 cm x 12.7 cm)
Peso del artículo 7.05 libras (aprox. 3.2 kg)
Componentes incluidos Stirrer apparatus (stir bar)

Garantía y soporte

The Corning PC-420D Stirring Hot Plate comes with a garantía de dos años from the date of purchase, covering defects in materials and workmanship under normal use. This warranty meets European Standards (EN).

For technical assistance, warranty claims, or service inquiries, please contact Corning customer support. Please have your model number (PC-420D or 6795-420D) and purchase date available when contacting support.

For the most up-to-date contact information, please visit the official Corning websitio o consulte la documentación incluida con su producto.

Documentos relacionados - PC-420D

Preview Manual de instrucciones de placas calefactoras, agitadores y agitadores/placas calefactoras Corning
Manual de instrucciones para placas calefactoras, agitadores y agitadores/placas calefactoras con pantalla digital Corning, así como para el controlador de temperatura 6795PR. Incluye instrucciones de funcionamiento, información de seguridad, especificaciones técnicas y detalles de la garantía.
Preview Manual de instrucciones de placas calefactoras, agitadores y agitadores/placas calefactoras Corning (120 V)
Manual de instrucciones completo para las placas calefactoras, agitadores y agitadores/placas calefactoras de 120 V de Corning (series PC-200, PC-400 y PC-600). Incluye información general, funcionamiento, mantenimiento, especificaciones y seguridad.
Preview Manual del usuario del sistema Corning Everon 6000 MIMO: configuraciones, LED y especificaciones de RF
Manual de usuario detallado de los sistemas MIMO Everon serie 6000 de Corning Optical Communications. Abarca configuraciones 2x2, 2TO1 y 4TO1 con componentes dLRU, DEU, DCU, RIU, BTS, dMRU y dHRU. Incluye definiciones de estado de LED, versiones de lanzamiento y advertencias sobre exposición crítica a radiofrecuencia para una instalación profesional.
Preview Manual del usuario de Corning MRU y HRU: Guía de instalación y funcionamiento
Este manual de usuario proporciona instrucciones completas para la instalación, configuración y funcionamiento de los sistemas MRU (Unidad remota de potencia media digital) y HRU (Unidad remota de alta potencia) de Corning, componentes esenciales para la infraestructura de red inalámbrica.
Preview Guía de selección de soportes permeables de Corning: insertos de cultivo celular Transwell y Falcon
A comprehensive guide from Corning and Fisher Scientific to select the optimal permeable supports, including Transwell and Falcon cell culture inserts, based on membrane type, pore size, format, and surface treatment for various research applications.
Preview Manual de instrucciones de la bioimpresora Corning Matribot
Manual de instrucciones completo para la bioimpresora Corning Matribot, que detalla la configuración, el funcionamiento, el mantenimiento y la resolución de problemas para aplicaciones de bioimpresión avanzadas en ingeniería de tejidos y cultivo de células en 3D.