Extech SDL150

Manual del usuario del medidor de oxígeno disuelto SD Logger Extech SDL150

Modelo: SDL150

1. Introducción

The Extech SDL150 is a Dissolved Oxygen Meter with an integrated SD card data logger. This device is designed for precise measurement of dissolved oxygen (DO) concentration and temperature in various applications. It records data directly onto an SD card in Excel format for easy transfer and analysis on a personal computer.

Key features include a dual backlit display, automatic temperature compensation, offset adjustment for relative measurements, and user-programmable samptasas de ling.

Extech SDL150 Dissolved Oxygen Meter with probe

Figura 1: Extech SDL150 Dissolved Oxygen Meter and Probe. This image shows the main unit with its digital display and control buttons, connected to the dissolved oxygen probe.

2. Contenido del paquete y configuración

2.1. Contenido del paquete

Verify that all items are present and in good condition upon unpacking. The standard package includes:

  • Extech SDL150 Dissolved Oxygen Meter
  • Dissolved Oxygen Probe with built-in temperature sensor
  • 6 pilas AA
  • SD card (2G capacity)
  • Two spare replacement membranes for the DO probe
  • Electrolyte solution for the DO probe
  • Estuche rígido

2.2. Configuración inicial

  1. Instalación de la batería: Open the battery compartment on the back of the meter and insert the 6 AA batteries, observing correct polarity.
  2. Inserción de tarjeta SD: Locate the SD card slot, typically on the side of the meter. Insert the provided 2G SD card until it clicks into place.
  3. Probe Preparation:
    • Unscrew the protective cap from the dissolved oxygen probe.
    • Fill the probe's membrane cap with the supplied electrolyte solution. Ensure no air bubbles are trapped.
    • Carefully screw the membrane cap back onto the probe, ensuring the membrane is intact and free of damage.
  4. Conexión de la sonda: Connect the prepared DO probe to the designated input port on the top of the SDL150 meter.
  5. Encendido: Presione el FUERZA para encender el dispositivo.

3. Instrucciones de funcionamiento

3.1. Medición básica

  1. Calibración: Before first use and periodically thereafter, calibrate the dissolved oxygen probe according to the instructions in the full user manual (not provided here). This typically involves air calibration or zero-oxygen solution calibration.
  2. Inmersión: Immerse the DO probe into the sample solution. Ensure the membrane is fully submerged and free from air bubbles.
  3. Pantalla de lectura: The dual backlit display will show the dissolved oxygen concentration (in mg/L or %) and the temperature (in °F or °C).
  4. Selección de unidades: Utilice el FUNCIÓN button to cycle through different measurement units (mg/L, %, °F, °C) if applicable.
Extech SDL150 Dissolved Oxygen Meter in use in a laboratory setting

Figura 2: Extech SDL150 in a laboratory environment. The meter is shown measuring dissolved oxygen in a beaker, demonstrating its application.

3.2. Registro de datos

  • Registro manual: Presione el REGISTRO button to manually store the current reading. The meter can store up to 99 manual readings.
  • Automatic Logging:
    • Access the settings menu (refer to the full manual for specific button combinations) to set the desired samptasa de ling.
    • Una vez que el sampling rate is set, activate automatic logging. The meter will record up to 20 million readings on the 2G SD card.
    • Data is stored in Excel-compatible format for easy transfer to a PC via the SD card.

3.3. Funciones adicionales

  • Retención de datos: Presione el SOSTENER Botón para congelar la lectura actual en pantalla. Presione nuevamente para liberar.
  • Mínimo máximo: Utilice el MÍNIMO MÁXIMO botón para view the minimum and maximum recorded values during a measurement session.
  • Iluminar desde el fondo: The display features a backlight for improved visibility in low-light conditions. Activate it using the dedicated backlight button (often indicated by a light bulb icon).
  • Apagado automático: The meter includes an auto power-off function to conserve battery life. This feature can typically be disabled or adjusted in the settings.
  • Ajuste de compensación: The offset adjustment function allows for zeroing the meter for relative measurements. Consult the full manual for detailed instructions on using this feature.

4. Mantenimiento

4.1. Probe Membrane and Electrolyte Replacement

The dissolved oxygen probe requires periodic maintenance, specifically the replacement of the membrane and electrolyte solution. This ensures accurate readings and extends the life of the probe.

  1. Unscrew the membrane cap from the probe body.
  2. Carefully remove the old membrane and discard it.
  3. Rinse the probe tip and the inside of the new membrane cap with distilled water.
  4. Fill the new membrane cap with fresh electrolyte solution, ensuring no air bubbles are present.
  5. Gently screw the new membrane cap onto the probe body. Avoid overtightening.
  6. Allow the probe to stabilize for a few minutes before use and perform a calibration.

4.2. Limpieza y almacenamiento

  • Limpieza del medidor: Limpie el exterior del medidor con un paño suave y húmedo.amp Paño. No utilice limpiadores abrasivos ni disolventes.
  • Limpieza de la sonda: After each use, rinse the probe tip with distilled water. Store the probe with its protective cap on, ideally with a small amount of electrolyte solution to keep the membrane moist, as specified in the full manual.
  • Almacenamiento a largo plazo: For extended storage, remove the batteries from the meter and ensure the probe is properly prepared as per the full manual's instructions. Store the unit and accessories in the hard carrying case in a cool, dry place.

5. Solución de problemas

This section provides general guidance for common issues. For detailed troubleshooting steps, refer to the comprehensive user manual provided with your Extech SDL150.

  • Sin energía: Verifique la instalación de las baterías y asegúrese de que no estén agotadas. Reemplácelas si es necesario.
  • Lecturas inexactas:
    • Realice una calibración.
    • Check the probe membrane for damage or air bubbles. Replace the membrane and electrolyte if needed.
    • Ensure the probe is properly connected to the meter.
  • La tarjeta SD no graba:
    • Verify the SD card is correctly inserted.
    • Check if the SD card is full. Transfer data to a PC and clear the card if necessary.
    • Ensure the logging function is activated and the sampling rate is set.

6. Especificaciones

ParámetroValor
Dissolved Oxygen Range0 to 20.0 mg/L (0 to 100.0%)
Rango de temperatura0 a 50 °C (32 a 122 °F)
Precisión de la medición+/-0.5%
Compensación de temperaturaAutomatic, 0 to 50°C
Capacidad de registro de datos99 manual readings, 20M readings via 2G SD card
Formato de datosExcel compatible (.xls)
Interfaz de PCBuilt-in (via SD card)
Fuente de poder6 pilas AA (incluidas)
Dimensiones (L x An x Al)7.2 x 2.9 x 1.9 pulgadas (18.3 x 7.4 x 4.8 cm)
Peso1.61 libras (0.73 kg)

7. Garantía y soporte

For information regarding warranty coverage, technical support, or service, please contact Extech customer service. Refer to the official Extech website or the documentation included with your product for current contact details and warranty terms.

Fabricante: Extec

Documentos relacionados - SDL150

Preview Catálogo completo de Extech: medidores ambientales, anemómetros y equipos de prueba
Explore la amplia gama de medidores ambientales, termoanemómetros, detectores de gas, comprobadores de calidad de agua, medidores de sonido, medidores de luz y otros instrumentos de prueba y medición de precisión de alta calidad de Extech diseñados para aplicaciones industriales, HVAC/R y profesionales.
Preview Manual del usuario del medidor de humedad, temperatura, flujo de aire y luz 4 en 1 EXTECH 45170
Manual de usuario del EXTECH 45170, un medidor 4 en 1 que mide humedad, temperatura, flujo de aire y luz. Incluye información sobre el funcionamiento, las especificaciones y la garantía.
Preview Manual del usuario del protector de radiación UV-AB Extech UV505
Manual del usuario completo para el transmisor de radiación UV-AB Extech UV505. Incluya instrucciones operativas, técnicas específicas, información de seguridad y mantenimiento.
Preview Manual del usuario del fotómetro UV-AB Extech UV505
Manual de usuario del fotómetro UV-AB Extech UV505. Este documento proporciona información detallada sobre las características, el funcionamiento, las especificaciones, las precauciones de seguridad y el mantenimiento del dispositivo para la medición de luz UVA y UVB.
Preview Extech EX810 1000 Amp Klemmeter con termómetro IR Gebruikershandleiding
Manipulación detallada para Extech EX810 1000 Amp Klemmeter con termómetro IR ingebouwde. Bevat instrucciónes, veiligheidsrichtlijnen y technische especificaties voor nauwkeurige metingen van spanning, stroom, weerstand, capacidad, frecuencia y temperatura.
Preview Guía del usuario de la serie EX650 de Extech: Cl digital de verdadero valor eficaz de 600 Aamp Metros
Guía de usuario completa para la serie EX650 de Extech de clavijeros digitales de 600 A con RMS verdaderoamp Medidores, que abarcan los modelos EX650 y EX655. Aprenda sobre el funcionamiento, las características, la seguridad y las especificaciones para obtener voltajes CA/CC precisos.tage, corriente, resistencia y más mediciones.