Modelo MSA 10178557

Manual del usuario del detector multigas MSA ALTAIR 4XR

Modelo: 10178557 | Marca: MSA

Introducción

The MSA ALTAIR 4XR Multigas Detector is a robust, handheld, portable gas monitor designed for critical safety applications. It is capable of simultaneously detecting up to four gases: Lower Explosive Limit (LEL), Oxygen (O2), Hydrogen Sulfide (H2S), and Carbon Monoxide (CO). Engineered for durability, this device features a tough over-molded rubber armor, is MIL-STD-810G drop protected, and boasts an IP68 rating for dust and water resistance, ensuring reliable performance in demanding environments.

MSA ALTAIR 4XR Multigas Detector front view

Figura 1: Frente view of the MSA ALTAIR 4XR Multigas Detector, showing the display and control buttons.

Características principales

Configuración y calibración

Before initial use or after prolonged storage, it is essential to perform a fresh air setup (zero calibration) and a span calibration to ensure accurate readings. Always perform these procedures in a clean, fresh air environment free from contaminants.

Equipo requerido:

Zero Calibration (Fresh Air Setup):

  1. Power on the ALTAIR 4XR detector by pressing and holding the middle button. Allow it to complete its startup sequence and warm-up procedure.
  2. When prompted for a "Fresh Air Setup" (FAS), select "No" to proceed to normal operating mode. This ensures a high-quality zero calibration.
  3. Once in normal operating mode, press and hold the right button until the display shows "ZERO CAL?".
  4. Press the middle button to confirm "Zero Cal". The unit will enter "SENSOR REFRESH" mode. Ensure the detector is in a clean air environment, free from any gas sources or residual gas.
  5. Allow the unit to complete the zero calibration. The display will indicate when it's complete.

Calibración de rango:

  1. Ensure the calibration gas cylinder has the correct gas mixture and is not expired. The standard MSA calibration gas mixture includes: Hydrogen Sulfide (20 ppm), Carbon Monoxide (60 ppm), Pentane Simulant (58% LEL), Oxygen (15.0%), and Nitrogen (Balance).
  2. Attach the 0.25 L/min Regulator to the calibration gas cylinder. Ensure the regulator's valve is open slightly to purge any residual air before fully tightening it onto the cylinder. This prevents moisture from entering the cylinder.
  3. Connect the tubing from the regulator to the calibration adapter.
  4. Attach the calibration adapter to the MSA ALTAIR 4XR detector. The adapter should securely cover the sensor inlets.
  5. From the normal operating mode, press and hold the right button until the display shows "SPAN CAL?".
  6. Press the middle button to confirm "Span Cal". The unit will prompt for gas application.
  7. Immediately turn on the gas flow from the cylinder by opening the regulator valve. The detector will begin to read the gas concentrations.
  8. Monitor the display. The values for each gas (LEL, O2, H2S, CO) should stabilize near the concentrations listed on your calibration gas cylinder. The unit will indicate "SPAN PASS" when calibration is successful.
  9. Once "SPAN PASS" is displayed, turn off the gas flow and remove the calibration adapter and tubing from the detector.
Worker using MSA ALTAIR 4XR Multigas Detector in a confined space

Figure 2: The MSA ALTAIR 4XR Multigas Detector in use, demonstrating its compact size and clear display for field applications.

Operación

After successful calibration, the ALTAIR 4XR is ready for use. The large, clear display shows real-time gas concentrations. The unit continuously monitors the environment and will activate audible, visual (LEDs), and vibrating alarms if gas levels exceed predefined thresholds or if a man-down event is detected via MotionAlert.

Mantenimiento

Regular maintenance is crucial for the longevity and accuracy of your MSA ALTAIR 4XR. Keep the device clean and free from debris. Periodically inspect the sensor inlets for blockages. The unit is compatible with MSA GALAXY GX2 Automated Test System for simplified and automated bump testing and calibration, which also helps in managing data and records.

Sensors are designed for long life but may require replacement over time. The device will provide an "end-of-sensor-life" warning. Replacement MSA XCell Sensors are available and can be easily installed in the plug-and-play slots.

Solución de problemas

If the device fails a bump test or calibration, first ensure that the calibration gas cylinder is not expired and contains the correct mixture. Check all connections for leaks. If sensor readings are erratic or consistently off, the sensor may be poisoned or at the end of its life and require replacement. For persistent issues, refer to the comprehensive troubleshooting guide in the full product manual or contact MSA support.

Presupuesto

EspecificaciónValor
MarcaMSA
EstiloO2, H2S, CO
Fuente de poderFunciona con pilas
ColorNEGRO
Dimensiones del producto1.37" de profundidad x 3" de ancho x 4.4" de alto
Peso del artículo0.51 libras (8.1 onzas)
AlarmaAudible, visual
Humedad de funcionamiento95 por ciento
Clasificación de temperatura superior50 grados centígrados
Tipo de sensorCatalytic, Electrochemical
Número de artículos1
FabricanteMSA
Número de pieza10178557
Número de modelo del artículo10178557
Fecha de primera disponibilidad25 de enero de 2018

Garantía y soporte

This product is covered by a 4-year warranty. For more information regarding warranty claims or technical support, please visit the official MSA website or contact their customer service. Ensure to register your product to activate the warranty and receive important updates.

For additional assistance, you may also refer to the comprehensive product manual provided with your device or contact a certified MSA service center.

Documentos relacionados - 10178557

Preview Manual de funcionamiento del detector multigas MSA ALTAIR 4XR
Manual de funcionamiento completo del detector multigas MSA ALTAIR 4XR, que abarca normas de seguridad, descripción del dispositivo, funcionamiento, mantenimiento, datos técnicos e información para realizar pedidos.
Preview Sistema de prueba automatizado MSA GALAXY GX2 para detectores de gas
El sistema de pruebas automatizadas MSA GALAXY GX2 ofrece pruebas y calibración sencillas e inteligentes para los detectores monogas MSA ALTAIR y ALTAIR PRO, así como para los detectores multigas ALTAIR 4XR y ALTAIR 5X. Cuenta con una interfaz intuitiva, pruebas automatizadas y funciones integrales de gestión de flotas.
Preview Manual de funcionamiento del detector multigas MSA ALTAIR 5X y 5XIR
Manual de funcionamiento completo para los detectores multigas MSA ALTAIR 5X y ALTAIR 5XIR, que cubre normas de seguridad, funcionamiento, mantenimiento, especificaciones técnicas y resolución de problemas para la detección de gases en entornos industriales.
Preview Sonda de bomba MSA ALTAIR: Detector de gas duradero para entornos peligrosos
Explore la sonda de bomba ALTAIR de MSA, un instrumento de detección de gases robusto y fiable. Conozca sus características principales, especificaciones técnicas, homologaciones e información de pedido para aplicaciones de seguridad industrial.
Preview Manual de funcionamiento del sistema de prueba automatizado MSA Galaxy GX2
Manual de funcionamiento del sistema de prueba automatizado MSA Galaxy GX2, que detalla la instalación, configuración, funcionamiento, mantenimiento y resolución de problemas para las pruebas del detector de gas.
Preview Manual de funcionamiento del detector de gas único MSA ALTAIR Pro - Guía de seguridad y uso
Este manual de funcionamiento proporciona instrucciones detalladas, pautas de seguridad, procedimientos de calibración y solución de problemas para el detector de gas único MSA ALTAIR Pro. Aprenda a usar y mantener eficazmente su instrumento de detección de gases.