APEX Domus 3

Manual del usuario del sistema de colchón de presión alterna Apex Medical Domus 3

Model: Domus 3

1. Introducción

This manual provides comprehensive instructions for the safe and effective use of the Apex Medical Domus 3 Alternating Pressure Mattress System. This system is designed to assist in the prevention and treatment of pressure injuries, ulcers, and bedsores for individuals requiring extended bed rest. Please read this manual thoroughly before setup and operation.

Apex Medical Domus 3 Alternating Pressure Mattress System with pump on a hospital bed.

Figura 1.1: The Apex Medical Domus 3 Alternating Pressure Mattress System. This image displays the complete system, including the mattress installed on a hospital bed frame and the external pump unit. The mattress features individual air cells, and the pump is shown connected to the mattress.

2. Información de seguridad

Adherence to the following safety guidelines is crucial for patient well-being and proper system function.

3. Componentes

The Apex Medical Domus 3 system typically includes the following components:

4. Instrucciones de configuración

  1. Prepare el marco de la cama: Place the mattress system directly onto a standard hospital or home bed frame. Ensure the bed frame is clean and free of sharp objects that could damage the mattress.
  2. Colocación del colchón: Unroll the Domus 3 mattress on the bed frame with the air hose connection end positioned towards the foot of the bed where the pump will be placed. Ensure the CPR strap is accessible.
  3. Conecte la bomba: Attach the air hoses from the mattress to the corresponding connectors on the variable pressure pump unit. Ensure connections are secure.
  4. Monte la bomba: Hang the pump unit securely at the foot of the bed using its integrated hangers.
  5. Conecte la electricidad: Plug the pump's power cord into a grounded electrical outlet.
  6. Inflación inicial: Turn on the pump. The mattress will begin to inflate. Allow approximately 20-30 minutes for the mattress to fully inflate and stabilize before placing a patient on it.
Apex Medical Domus 3 mattress cover with zipper and water-resistant material detail.

Figura 4.1: Mattress Cover Features. An image showcasing two features of the mattress cover: a close-up of the zipper-all-around design for easy removal and replacement, and a magnified view of the Nylon/PU cover material demonstrating its water-resistant properties with water droplets on the surface.

5. Instrucciones de funcionamiento

5.1 Encendido y apagado

5.2 Modos de terapia

The Domus 3 offers two therapy modes:

Refer to the pump unit's controls to select the desired therapy mode.

Diagram illustrating the alternating pressure redistribution cycle of the mattress.

Figura 5.1: Alternating Pressure Cycle. This diagram shows a human figure lying on the mattress, with an illustration of the alternating pressure cycle. Blue blocks represent inflated air cells, and arrows indicate the cyclical inflation and deflation pattern designed to redistribute pressure and prevent injury.

5.3 Pressure Adjustment (Firmness)

The pump unit features a variable pressure setting with a weight indicator.

5.4 Low Air Loss Function

Air cells 3 through 11 are designed with a low air loss feature. This function continuously ventilates air to wick moisture and heat away from the mattress surface, contributing to patient comfort and skin integrity. This operates automatically when the system is powered on.

Close-up of Apex Medical Domus 3 mattress showing low air loss ventilation.

Figura 5.2: Low Air Loss Feature. A detailed view of the mattress surface, highlighting the low air loss feature. Blue arrows indicate airflow through the air cells, designed to reduce moisture and heat for patient comfort and prevent skin maceration.

6. Mantenimiento y cuidado

6.1 Cleaning the Mattress and Cover

6.2 Air Cell Replacement

Each of the 20 individual 8-inch air cells can be removed and replaced if damaged.

  1. Turn off the pump and disconnect it from the mattress.
  2. Unzip the mattress cover.
  3. Carefully detach the damaged air cell from its connections within the mattress structure.
  4. Install a new, compatible air cell, ensuring it is securely connected.
  5. Zip the mattress cover back into place and re-connect the pump for inflation.

7. Solución de problemas

ProblemaPosible causaSolución
Mattress not inflating or losing pressure
  • Cable de alimentación desconectado
  • Air hoses not securely connected
  • Leak in mattress or air cells
  • Mal funcionamiento de la bomba
  • Check power connection to outlet and pump.
  • Ensure air hoses are firmly attached to both mattress and pump.
  • Inspect mattress and individual air cells for punctures or damage. Replace damaged cells if necessary.
  • Contact customer support if pump appears to be malfunctioning.
Low pressure alarm sounds
  • Mattress not fully inflated
  • Fuga de aire
  • Patient weight exceeds capacity
  • Allow more time for inflation.
  • Check for leaks in mattress or hose connections.
  • Verify patient weight is within the 440 lbs capacity.
Pump is unusually loud
  • Obstruction around pump fan
  • Mal funcionamiento de la bomba
  • Ensure adequate ventilation around the pump unit.
  • Contact customer support if noise persists. (Note: Pump noise level is typically ≤ 30dB(A)).

8. Especificaciones

CaracterísticaDetalle
ModeloCasa 3
Mattress Dimensions (L x W x H)78.7 x 35.4 x 8 pulgadas
Peso del sistemaAproximadamente 22.3 libras
Capacidad de peso440 libras
Tipo de telaNylon, Polyurethane (Cover)
Características especialesLow Air Loss, Alternating Pressure, Water Resistant Cover, CPR Strap, Low Pressure Alarm
Nivel de ruido de la bomba≤ 30dB (A)
FabricanteApex Medical
Dimensions of Apex Medical Domus 3 mattress and pump system.

Figura 8.1: System Dimensions. A visual representation of the dimensions for both the Apex Medical Domus 3 mattress and its accompanying pump unit. Key measurements for length, width, and height are indicated for both components.

Apex Medical Domus mattress showing two height options, 5-inch and 8-inch.

Figura 8.2: Mattress Height Options. This image illustrates the two available height options for the Domus mattress: a 5-inch version and an 8-inch version. The current product is the 8-inch model, providing enhanced support.

9. Garantía y soporte

9.1 Información de garantía

The Apex Medical Domus 3 Alternating Pressure Mattress System comes with a Garantía de 1 años on both the mattress and the pump unit. This warranty covers manufacturing defects and malfunctions under normal use. Please retain your proof of purchase for warranty claims.

9.2 Atención al cliente

For technical assistance, troubleshooting beyond this manual, or warranty inquiries, please contact Apex Medical customer support. Refer to the product packaging or the official Apex Medical webSitio para obtener información de contacto actualizada.

Documentos relacionados - Casa 3

Preview Guía rápida del modo de proveedor automático de APEX XT
Una guía rápida para profesionales médicos sobre cómo utilizar el dispositivo APEX XT Auto en modo de proveedor, incluida la configuración, la navegación y el restablecimiento de los medidores de cumplimiento.
Preview Manual del usuario de la bomba APEX VAC: Instrucciones y guía de seguridad
Manual de usuario completo para la bomba APEX VAC, que detalla el funcionamiento, la seguridad, las especificaciones técnicas y la solución de problemas para los modelos VacPro, VacMaxi y VacPlus de Apex Medical Corp.
Preview Manual de instrucciones del sistema CPAP automático APEX XT
Manual de instrucciones completo para el sistema CPAP automático APEX XT (modelo 9S-005720), que detalla la configuración, el funcionamiento, el mantenimiento, la solución de problemas y las especificaciones técnicas para el tratamiento de la apnea obstructiva del sueño (AOS).
Preview Manual de instrucciones del sillón de masaje APEX-V2: Guía de instalación, funcionamiento y mantenimiento
Manual de instrucciones completo del sillón de masaje APEX-V2, que abarca precauciones de seguridad, instalación, funcionamiento, configuración, especificaciones, solución de problemas, limpieza y mantenimiento. Aprenda a configurar y usar su sillón de masaje APEX-V2 eficazmente.
Preview Manual del propietario del sistema de seguridad DESTINY
Este manual del propietario del sistema de seguridad Apex DESTINY proporciona orientación completa sobre la instalación, el funcionamiento, los modos de armado (AUSENTE, EN CASA, NOCHE), acceso remoto, funciones de derivación y funciones opcionales de automatización del hogar.
Preview Manual de instrucciones del sillón de masaje Apex AP-Pomp
Manual de usuario completo para el sillón de masaje Apex AP-Pomp, que abarca precauciones de seguridad, mantenimiento, instalación, funciones del controlador, solución de problemas y especificaciones.