OFITE API Filter Press with CO2 Pressure Assembly

Manual del usuario del conjunto de filtro prensa API con presión de CO2

Model: API Filter Press with CO2 Pressure Assembly

Marca: OFITE

1. Introducción

The OFITE API Filter Press with CO2 Pressure Assembly is a precision instrument designed for evaluating the filtration properties of drilling fluids. This device provides a quick and easy method to determine fluid loss and filter cake characteristics under controlled pressure conditions. It is engineered to conform to American Petroleum Institute (API) specifications, making it suitable for both field and laboratory applications in the oil and gas industry.

Understanding the filtration behavior of drilling fluids is crucial for optimizing drilling operations, preventing formation damage, and ensuring wellbore stability. This manual provides comprehensive instructions for the safe and effective use, maintenance, and troubleshooting of your API Filter Press.

2. Información de seguridad

Siga siempre las siguientes pautas de seguridad para evitar lesiones y daños al equipo:

3. Componentes sobreview

OFITE API Filter Press with CO2 Pressure Assembly
Figura 3.1: OFITE API Filter Press with CO2 Pressure Assembly. This image shows the complete filter press unit, including the pressure cell, frame, CO2 pressure regulator with gauge, and the graduated cylinder for collecting filtrate. A box of filter paper is visible at the base.

4. Instrucciones de configuración

  1. Desembalaje e Inspección: Carefully unpack all components. Inspect for any shipping damage. Report any damage to the supplier immediately.
  2. Asamblea:
    • Place the filter press frame on a stable, level surface.
    • Attach the CO2 pressure regulator assembly to the pressure cell inlet, ensuring a tight seal.
    • Connect the CO2 pressure regulator to a suitable CO2 cylinder. Ensure the cylinder is secured to prevent tipping.
  3. Preparation of Pressure Cell:
    • Ensure the pressure cell is clean and dry.
    • Place a new filter paper (e.g., OFITE Catalog No. 140-55) at the bottom of the pressure cell, ensuring it lies flat and covers the screen.
    • Secure the bottom cap of the pressure cell, ensuring the O-ring is properly seated for a leak-proof seal.
  4. Fluido Sample Preparación: Preparar el fluido de perforaciónample according to relevant API recommended practices or your specific testing protocol.

5. Instrucciones de funcionamiento

Follow these steps for conducting a filtration test:

  1. Carga Sampen: Pour the prepared drilling fluid sample into the pressure cell. Do not overfill.
  2. Seal Cell: Securely attach the top cap of the pressure cell, ensuring the O-ring is properly seated.
  3. Position Graduated Cylinder: Place a clean, dry graduated cylinder beneath the filtrate outlet of the pressure cell.
  4. Aplicar presión:
    • Slowly open the valve on the CO2 cylinder.
    • Adjust the pressure regulator to the desired test pressure (e.g., 100 psi for standard API tests). Monitor the gauge carefully.
  5. Iniciar prueba: Once the desired pressure is reached, immediately start a timer.
  6. Monitor Filtrate: Observe the volume of filtrate collected in the graduated cylinder over the specified test duration (e.g., 30 minutes for standard API tests).
  7. Resultados del récord: At the end of the test duration, record the total volume of filtrate collected.
  8. Liberar presión: Slowly close the CO2 cylinder valve, then carefully open the pressure relief valve on the regulator to depressurize the system. Ensure the gauge reads zero before proceeding.
  9. Desmontar y limpiar: Carefully disassemble the pressure cell. Remove the filter paper and fluid sample. Proceed to cleaning as described in the Maintenance section.

6. Mantenimiento

Regular maintenance ensures the longevity and accuracy of your API Filter Press.

6.1. Limpieza

6.2. Inspección y reemplazo de la junta tórica

6.3. Pressure Regulator Care

7. Solución de problemas

ProblemaPosible causaSolución
Pressure loss during testLeaking O-rings; Loose connections; Damaged pressure cell.Inspect and replace O-rings. Tighten all connections. Inspect pressure cell for cracks or damage; replace if necessary.
No filtrate collectedClogged filter paper; Insufficient pressure; Blocked filtrate outlet.Ensure fresh filter paper is used. Verify pressure setting. Clean filtrate outlet.
Resultados inconsistentesimpropioample preparation; Varying pressure; Contaminated equipment.Ensure consistent sample preparation. Verify pressure regulator accuracy. Thoroughly clean all components before each test.
El manómetro no lee correctamenteGauge malfunction; Regulator issue.Contact OFITE support for calibration or replacement of the pressure regulator/gauge.

8. Especificaciones

9. Garantía y soporte

OFITE products are manufactured to high standards of quality and reliability. For specific warranty terms and conditions, please refer to the documentation provided with your purchase or visit the official OFITE websitio.

For technical support, spare parts, or service inquiries, please contact OFITE directly:

Al contactar con el soporte técnico, tenga a mano el modelo del producto y ASIN (B09CVBKG5L) readily available.

Documentos relacionados - API Filter Press with CO2 Pressure Assembly

Preview Manual de instrucciones del filtro prensa de media área OFITE 140-60
Manual de instrucciones del filtro prensa completo de media área OFITE 140-60, que detalla sus componentes, funcionamiento, mantenimiento e información sobre la garantía. Diseñado para el análisis de filtración a baja presión de fluidos de perforación.
Preview Manual de instrucciones del filtro prensa OFITE HTHP - Modelos 170-00-3 y 170-01-3
This comprehensive instruction manual details the operation, maintenance, and safety procedures for the OFITE High Temperature High Pressure (HTHP) Filter Press, including models 170-00-3 (115 V) and 170-01-3 (220 V). It covers specifications, components, testing procedures, and warranty information for drilling fluid analysis.
Preview OFITE 4-Unit HTHP Filter Press Instruction Manual
Instruction manual for the OFITE 4-Unit High-Temperature High-Pressure (HTHP) Filter Press, detailing setup, operation, safety procedures, load cell option, software setup, calibration, driver installation, temperature controller usage, and warranty information.
Preview Manual de instrucciones del filtro prensa OFITE HTHP con celdas roscadas
Manual de instrucciones del filtro prensa OFITE de alta temperatura y alta presión (HTHP) con celdas roscadas (modelos 171-181-N y 171-181-1-N). Detalla la evaluación de fluidos de perforación, lechadas de cemento, etc., a temperaturas y presiones elevadas, cumpliendo con las especificaciones API. Abarca la operación, la seguridad y el mantenimiento.
Preview Filtro prensa OFITE HTHP para pruebas de cemento - Manual de instrucciones
Manual de instrucciones del filtro prensa OFITE HTHP (modelo 170-182) diseñado para evaluar las características de filtración de fluidos de perforación, lechadas de cemento y otros fluidos en condiciones de alta temperatura y presión.
Preview Filtro prensa OFITE HTHP con celdas roscadas para pruebas de cemento - Manual de instrucciones
Comprehensive instruction manual for the OFITE HTHP Filter Press with Threaded Cells (Model #171-192), designed for cement testing under high temperature and pressure conditions. Learn about specifications, operation, maintenance, and safety.